FEATURES
- Traceable to SI 1
- Based on NIST Proven Two-Pressure Principle
- Push Button (Keypad) Control
- Automated Control of User Setpoints
- 3900 ControLog® Automation Software
- HumiCalc® with Uncertainty Software
- Automatically Applies Enhancement Factors
- Computerized Internal Transducer Calibration
- RS-232C Serial Interface
- Only Nine Square Feet of Floor Space
- Timed/Formatted Output to Printer
- Battery Backed-up Real Time Clock
- Backlit Liquid Crystal Display
1 Traceable to the International System of Units (SI) through NIST-maintained standards.
SPECIFICATIONS 2
Frost / Dew Point Range: | –95 to +10 °C |
Frost / Dew Point Uncertainty: 2 | (–95 to –90 °C) 0.9 °C |
(–90 to –80 °C) 0.5 °C |
(–80 to –70 °C) 0.2 °C |
(–70 to 10 °C) 0.1 °C |
Parts Per Million Range: | 0.05 to 12000 PPMv |
Relative Humidity Range | 0.0002 to ~50% |
Saturation Pressure Range: | Ambient to 300 psiA |
Saturation Pressure Uncertainty (10-50 psiA): 2 | 0.05 psiA |
Saturation Pressure Uncertainty (50-300 psiA): 2 | 0.30 psiA |
Saturation Temperature Range: | –80 to +12 °C |
Saturation Temperature Uncertainty: 2 | 0.08 °C |
Saturation Temperature Heating/Cooling Rate: | 2 Minutes Per °C Average |
Test Pressure Range: | Ambient to 50 psiA |
Test Pressure Uncertainty (10 - 50 psiA): 2 | 0.05 psiA |
Test Pressure Range (Option): | Ambient to 150 psiA |
Test Pressure Uncertainty (10 - 150 psiA)(Option): | 0.05 psiA |
Test Temperature Range: | 0 to 50 °C |
Test Temperature Uncertainty: 2 | 0.08 °C |
Display Resolution: | 0.01 |
Gas Flow Rate Range: | 0.1 to 5 L/min |
Gas Flow Rate Range (Option): 3 | 0.5 to 10 L/min |
Gas Flow Rate Resolution: | 0.02 L/min |
Gas Flow Rate Uncertainty: 2 | 0.2 L/min |
Gas Type: | Air or Nitrogen |
Gas Pressure Rating (MAWP): | 350 psiG |
Refrigeration: | 1/3 HP R-134A & 1/3 HP R-23 in Cascade |
Heating: | Stainless Steel Immersion Heaters |
Test Port: | 1/4 Inch Swagelok® Tube Fitting (6.35mm) |
Physical Dimensions: | 37.5" H x 23" W x 30" D (953mm x 584mm x 762mm) |
2 Represents an expanded uncertainty using a coverage factor, k=2, at an approximate level of confidence of 95%.
3 Gas Supply (Option): 350 psiG @ 0.35 cfm (10 L/min) w/ambient pressure FP <–80 °C.
UTILITIES
Electrical Power: | 200-230/210-240 V~, 10 A, 50/60 Hz |
Gas Supply (External): | 350 psiG, 5 L/min, with ambient pressure frost point <–80 °C |
Floor Space: | 9 ft 2 (0.84 m 2) |
ENVIRONMENTAL
Operating Temperature: | 15 to 30 °C |
Storage Temperature: | 0 to 50 °C |
Humidity: | 5 to 95% Non-condensing |
PRINCIPLE OF OPERATION
The “two-pressure two-temperature” generation process involves saturating a continuous stream of air or nitrogen with water vapor at a known temperature and pressure. The saturated high pressure air then passes through an expansion valve where it expands to a lower pressure. The 3900 generates a particular humidity by first selecting a suitable saturation temperature, Ts. It then determines the saturation pressure, Ps, required to establish the correct saturation vapor pressure. The precision of the system is determined by the accuracy of the temperature and pressure measurements and on the constancy of them throughout. When setpoint equilibration has been reached, the indication of saturation temperature, saturation pressure, test temperature, and test pressure may be used in the determination of all hygrometric parameters. Furthermore, because the humidity generated is based solely on the fundamental principles of temperature and pressure, no humidity sensing is used to measure or control the amount of water vapor produced by this system. The 3900 operates using an on board multifunction CPU in conjunction with other peripheral cards to perform calculation and control functions. The embedded computer control system allows the 3900 to generate known humidity levels completely unattended with visual indications of system status displayed in real time on the Liquid Crystal Display.

Elemental Schematic of the 3900 generator.
COMPUTER CONTROL

3900 Control Display Screen
Control / Display Screen
This frees the operating technician from the task of system monitoring and adjustment. A computer and/or printer may be connected via the bi-directional RS-232C interface ports allowing remote setpoint control and continuous system data retrieval.
Temperature Control: Temperature setpoint control is attained by controlling the temperature of a circulating fluid medium that jackets the saturator of the generator. The saturation temperature is governed by the temperature of this medium, which is digitally controlled by the computer at any value between –80 °C and 12 °C through the use of PID (proportional-integral-derivative) algorithms.
Pressure and Flow Control: Pressure control and mass flow rate control are accomplished through computer actuation of electromechanical valve assemblies. Saturation pressure and mass flow are measured continuously and controlled using PID algorithms similar to those employed in temperature control.
Two-Pressure Two-Temperature Generator: Regulated compressed air or nitrogen is directed through the saturator, which is a fluid encapsulated heat exchanger containing several planes of pure ice or water. The saturator is maintained at the required saturation temperature and saturation pressure. As the gas thermally equilibrates, it becomes saturated with water vapor. The saturation temperature, Ts, and saturation pressure, Ps are measured at the point of final saturation. The saturation pressure is then reduced to test pressure, Pt, and the conditioned gas is admitted to the device under test (DUT) at the desired humidity conditions. The final pressure, Pt, and temperature, Tt, of the gas is measured within or just after the DUT. The DUT is then exhausted to atmosphere or to a back pressure regulator to achieve pressure control.
Calibration Display Screen

Calibration Display Screen
Calibration: Proper calibration of the temperature and pressure transducers ultimately determines the accuracy of the generator. This system employs an integral programmatic calibration scheme allowing the transducers to be calibrated while they are electrically connected to the humidity generator.This approach helps eliminate systemic errors that might be induced by removing the transducers from the generator. All calibration is performed mathematically by the computer so manual adjustments are not needed. Coefficients for each transducer are calculated by the computer and stored in the system’s nonvolatile memory until the next calibration is performed.

3900 Front Control Panel
The main panel has easy access to the test temperature and the test pressure probe connectors, and you can see your results promptly on the liquid crystal display.
APPLICATIONS
Virtually any humidity may be generated, for any length of time, within the operational limits of the generator. The output of the device under test may then be compared with the generator’s printed data for analysis.
Chilled Mirror Hygrometers: Connect the generator output to your chilled mirror hygrometer and you can: verify mirror temperature measurement accuracy (calibration) when the hygrometer is in thermal equilibrium with its environment; perform operational checks of the heatpump and optical components, before and after mirror cleaning and balancing; determine whether the hygrometer is controlling the mirror deposit in the liquid phase or ice phase when operating at dew and frost points below 0 °C; determine if the hygrometer is correctly calculating other humidity parameters; determine hygrometer repeatability, stability, and drift characteristics.
Humidity Sensors and Electrolytic Hygrometers: Connect the generator output to your Electrolytic Hygrometer, sampling system, special fixtures, or sensors and you can: determine humidity calibration accuracy and/or characterize humidity sensitivity by subjecting the humidity sensor to a variety of humidity levels; perform operational checks such as the sensing systems capability to correctly calculate and display other humidity parameters; determine repeatability, stability, hysteresis, and drift characteristics of various humidity sensing systems.
Environmental Testing: The 3900 can serve as a test bed for evaluation and R&D of humidity sensors, humidity sensing systems, and humidity sensitive products, e.g., polymers, composites, film, magnetic medium, pharmaceuticals, soil hydrology, consumables, electronics, optics, etc.
REFERENCE DOCUMENTS
Model 3900 Low Humidity Generation System Documents
Series 3900 operation and maintenance manual for the two-pressure two-temperature low humidity generator.
Model 3900 Uncertainty Analysis
This document describes a complete uncertainty analysis of the Model 3900 in accordance with NIST Technical Note 1297. Each error source is identified, quantified, and combined to form an expanded uncertainty with a coverage factor of 2.